ArgoCap – Feasibility and Safety of a Novel Over-the-scope Device to Facilitate Endoscopic APC Treatment
Date de publication : 2023
Éditeur : Universität
Nombre de pages : Non disponible
Résumé du livre
Abstract: Introduction
Argon plasma coagulation (APC) is an electrosurgical procedure used, among other indications, for treatment of dysplastic Barrett's mucosa. Homogeneous and safe application can be compromised by varying distances and suboptimal angle of the probe to the tissue. In this study, we present ArgoCap, a novel endoscopic device developed to facilitate endoluminal APC treatment. Objectives of this preclinical study were to assess feasibility and safety and to determine suitable APC settings.
Material and methods
One-hundred and thirty-two APC treatments of predefined areas using various APC settings were performed ex vivo in the opened porcine esophagus. Depth of thermal injury was assessed histologically. Feasibility of APC treatment in different locations was examined in 20 explanted porcine esophagi and in first in vivo porcine applications.
Results
APC treatment in all quadrants of the esophagus was feasible. Histologically, thermal effects involving the whole thickness of the mucosa were visible with all settings. APC with pulsed mode resulted in deep thermal damage with all power settings. No lesions of the muscular layer occurred using precise (E8, E9) and forced (10 W, 20 W) mode.
Conclusions
Esophageal APC using ArgoCap is feasible and safe. The device has the potential to improve APC treatment of larger mucosal areas